A Professional Guide to Using the Winpower Handheld Analyzer

Standard Operating Procedure: Performance Verification Using the Winpower Handheld Analyzer

Purpose and Scope

The integrity of all service documentation is dependent on the accuracy of performance data. This document provides the standard operating procedure for using the Winpower handheld oxygen analyzer to verify the performance of medical oxygen concentrators. The procedure covers the measurement of three key performance indicators: oxygen concentration, output flow, and system pressure. Adherence to this protocol is mandatory for ensuring that all service reports are built on a foundation of verifiable and trustworthy information.

Equipment Overview

The Winpower handheld analyzer is a purpose-built instrument for service professionals. Its measurement capabilities are:

  • Oxygen Concentration: 21% to 95.6%
  • Output Flow: 0 to 10 L/min
  • System Pressure

The device features a rechargeable internal battery and a simple three-button interface for power, backlight, and mode selection.

Pre-Test Protocol

Correct usage begins before the device is connected to a concentrator. The following preparatory steps are required to ensure data accuracy and prevent equipment damage.

  • Battery and Charging: To maintain battery health and device compatibility, the analyzer must be charged using the supplied 12-watt charger.
  • Concentrator Preparation: It is a critical procedural step to remove any humidifier bottle from the oxygen concentrator being tested. This action prevents moisture from entering the analyzer and compromising its internal sensors.

Measurement Procedure

  1. Power On and Stabilize: Power on both the Winpower analyzer and the oxygen concentrator. A brief warm-up period is required for systems to stabilize.
    • Allow approximately 10 seconds for the analyzer.
    • Allow up to 5 minutes for the oxygen concentrator. This ensures that collected data reflects the machine’s steady operational state.
  2. Connect Equipment: Securely connect the analyzer’s inlet port to the oxygen outlet of the concentrator.
  3. Capture Data: Using the mode selection button, cycle through the display screens to capture readings for oxygen concentration (%), flow (L/min), and pressure.
  4. Document Results: Record the stabilized measurements in the official service report. Following this disciplined protocol ensures that all performance verifications are definitive.

Video Reference

To support professional practice, a clear instructional video details the entire process, from unboxing to capturing accurate readings.

Watch the full instructional video here:

Technical Service Disclaimer

The content provided on this blog, including all articles, guides, and embedded videos, is for informational and educational purposes only. This information is intended to supplement, not replace, the official service manuals, user guides, and other documentation provided by the medical device manufacturer.

All maintenance, troubleshooting, and repair procedures on medical equipment must be performed in strict accordance with the manufacturer’s official service manual for the specific device model and serial number. Before attempting any procedure described or demonstrated on this blog, you are required to locate, read, understand, and adhere to all instructions, warnings, and safety precautions outlined in the official documentation.

The procedures described are intended for use by qualified, trained, and certified biomedical engineering technicians and professionals. Any action you take upon the information on this website is strictly at your own risk. The author and publisher will not be liable for any personal injury, damage to equipment, voided warranties, or other losses that may result from the use of this content.

No warranty, express or implied, is made regarding the accuracy, completeness, or reliability of the information provided. Always prioritize patient and personal safety by following manufacturer guidelines and your facility’s established safety and infection control protocols.