Site icon Oxygen Alliance

A Professional Guide to Using the Winpower Handheld Analyzer

Operation of a winpower handheld oxygen analyzer

Operation of a winpower handheld oxygen analyzer

Standard Operating Procedure: Performance Verification Using the Winpower Handheld Analyzer

Purpose and Scope

The integrity of all service documentation is dependent on the accuracy of performance data. This document provides the standard operating procedure for using the Winpower handheld oxygen analyzer to verify the performance of medical oxygen concentrators. The procedure covers the measurement of three key performance indicators: oxygen concentration, output flow, and system pressure. Adherence to this protocol is mandatory for ensuring that all service reports are built on a foundation of verifiable and trustworthy information.

Equipment Overview

The Winpower handheld analyzer is a purpose-built instrument for service professionals. Its measurement capabilities are:

The device features a rechargeable internal battery and a simple three-button interface for power, backlight, and mode selection.

Pre-Test Protocol

Correct usage begins before the device is connected to a concentrator. The following preparatory steps are required to ensure data accuracy and prevent equipment damage.

Measurement Procedure

  1. Power On and Stabilize: Power on both the Winpower analyzer and the oxygen concentrator. A brief warm-up period is required for systems to stabilize.
    • Allow approximately 10 seconds for the analyzer.
    • Allow up to 5 minutes for the oxygen concentrator. This ensures that collected data reflects the machine’s steady operational state.
  2. Connect Equipment: Securely connect the analyzer’s inlet port to the oxygen outlet of the concentrator.
  3. Capture Data: Using the mode selection button, cycle through the display screens to capture readings for oxygen concentration (%), flow (L/min), and pressure.
  4. Document Results: Record the stabilized measurements in the official service report. Following this disciplined protocol ensures that all performance verifications are definitive.

Video Reference

To support professional practice, a clear instructional video details the entire process, from unboxing to capturing accurate readings.

Watch the full instructional video here:

Exit mobile version