Site icon Oxygen Alliance

A Service Protocol for the Diamedica Oxygen Reservoir System

Diamedica oxygen reservoir system teardown

Diamedica oxygen reservoir system teardown

System Purpose and Scope

In any surgical environment, the continuous availability of medical oxygen is fundamental. The Diamedica oxygen reservoir system is engineered to provide a robust backup solution for anesthesia machines, such as the Diamedica Glina, when the primary power source is unstable or fails completely. It acts as a crucial buffer, storing hours of life-sustaining oxygen to ensure patient care continues without interruption. This protocol details the system’s components, safety features, and operating procedures for biomedical professionals responsible for safeguarding the anesthetic environment.

System Components and Specifications

The system is composed of two primary units: the oxygen reservoir vessel and the air pump.

Oxygen Reservoir Vessel
Air Pump Unit

System Safety Features

The pump’s pneumatic path and control systems are designed with multiple safety features critical for handling medical oxygen.

Operating Procedure: Reservoir Filling

The filling process is straightforward and automated.

  1. Connect the output of one or more oxygen concentrators to the air pump’s designated inlet port.
  2. Activate the system. The pump will begin pressurizing the reservoir.
  3. The pump will run until the reservoir is filled to capacity, at which point the integrated pressure switch will automatically shut off the compressor.
  4. The reservoir is now fully charged and ready to act as an independent oxygen source.

Maintenance Overview

Mastering the service procedures for the air pump is essential for maintaining the system at peak readiness. A complete disassembly, as demonstrated in the video guide, involves disconnecting the battery and pneumatic tubings and removing the compressor itself for inspection or service. This intimate knowledge allows for precise troubleshooting and ensures every component is functioning correctly, guaranteeing this critical backup system will perform flawlessly when needed most.

Video Reference

A detailed video presentation is available to provide a comprehensive visual walkthrough of the entire system. It covers the function of every component, a full teardown of the air pump, and a step-by-step demonstration of the filling procedure.

Watch the full instructional video here: